技术参数
取样方法 红外线试验法 (NDIR), 镀金光学, (with Telaire’s patented ABC Logic self calibration algorithm).
测量范围 0…2,000 ppm 工厂校准,O utput scalable using UIP up to 3,000 ppm with resulting decrease in
sensor accuracy
精确性 +/- 100 ppm @22℃ (72°F ) when compared with a factory reference
温度依存性 0.2% FS per℃
稳定性 < 2% of FS over life of sensor (15 years typical)
非线性 < 1% of FS
压力依存性 0.13% of reading per mm Hg
校准间隔 Not required *
响应时间 TBD, pending verification tests.
Estimated at 3…5 minutes for 90% step change
启动时间 < 2 minutes (operational)
10 minutes (maximum accuracy)
操作条件 0…50℃ (32…122°F ) 0…95% RH, non condensing
贮藏条件 -20…70℃ (-4…158°F )
0…95% RH, non condensing
温度 NTC 20k Thermistor
输出 Analog: 0…10V (100 Ohms output impedance)
Digital: RS-232 (TTL Level) communication
with Telaire UIP software
连接 Screw terminals for 18 28 AWG
电源要求 18…30 VAC RMS, 50/60Hz or 18…42 VDC, polarity protected
功率 1.75 VA average, 3.25 VA peak
AGENCY/CERTIFICA-TIONS FCC Part 15 Class B, CE, ROHS